Sterilization Q&A

  • Asked March 22nd, 2009

    Yes, the current ANSI/AAMI/ISO 11137-2:2006 includes VDmax dose substantiation for both 25 kGy and 15 kGy. However, if you are using VDmax 15 kGy, bioburden is limited to no greater than 1.5 CFU. In addition, there is AAMI TIR 33:2005, Sterilization of health care products – Radiation- Substantiation of a selected sterilization dose – Method VDmax, which expands the VDmax method to several other doses, 15 kGy up to 35 kGy at approximately 2 kGy increments, where bioburden range is defined for each dose. AAMI TIR 33:2005 is available for purchase at www.aami.org.

    .